Model 192

Ludlum Model 192, microR survey meter

Manufacturer
Ludlum Measurements
Reference
Model 192

Specifications

Product
Ludlum Model 192, microR survey meter
Ludlum part number
48-2945
Application
Low-level (microR) gamma survey
Detector (D x L)
NaI(Tl) scintillator, 5.1 x 2.5 cm (2.0 x 1.0 in.)
Detection Range
0 to 50 µSv/h (0 to 5000 µR/hr)
Sensitivity ( 137 Cs gamma)
Typically 650 cpm per µR/hr
Energy Response
Energy Dependent
Linearity
Reading within 10% of true value
Meter Dial
0 to 5 µR/hr, BAT TEST (Meter Faces tab lists other available meter faces)
Meter
6.4 cm (2.5 in.) arc, 1 mA analog type
Controls
Rotary Selector: Switches between instrument off, battery check, ranges: x1, x10, x100, x1000 Response Switch: Toggles between FAST (4 s) or SLOW (22 s) from 10% to 90% of final reading Reset Pushbutton: To zero meter, a
Alarms
This instrument incorporates dual action, latching type alarms: A fixed alarm point that can be set at any point from 10% of full scale to full scale, and is indicated by a constant audible tone, and the lamp turning on.
Battery Life
Typically 600 hours with alkaline batteries (battery condition can be checked on meter)
Size (H x W x L)
21.6 x 8.9 x 22.4 cm (8.5 x 3.5 x 8.8 in.) with handle
Weight
2.3 kg (5 lb), including batteries
Meter Face
Image (click to view larger)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line