Model 9-4

Ludlum Model 9-4, ion chamber survey meter

Manufacturer
Ludlum Measurements
Reference
Model 9-4

Specifications

Product
Ludlum Model 9-4, ion chamber survey meter
Ludlum part number
48-3739
Application
Gamma and X-ray exposure survey meter with beta response
Detection Range
0 to 500 mSv/h (0 to 50,000 mR/hr)
Energy Response
Within 20% of true value from 40 keV to 2 MeV
Linearity
Reading within 10% of true value
Window
7 mg/cm 2 metalized polyester; with slide open, allows gamma detection to 6 keV
Window Area
40 cm 2 (6.2 in 2 )
Chamber Volume
220 cm 3 (13.4 in 3 )
Chamber Construction
Carbon coated acrylic
Side Wall
1000 mg/cm 2 aluminum and acrylic
Beta Shield
Retractable 1000 mg/cm 2 phenolic slide with side button control
Beta Response
Factor of 4.8 difference between window open and closed measurements when exposed to a uranium slab
Meter Dial
0 to 5 mR/hr, BAT TEST
Meter
6.4 cm (2.5 in.) arc, 1 mA, pivot-and-jewel suspension
Controls
Range Selection: x1, x10, x100, x1k, x10k, plus instrument off Reset: The RESET circuit grounds the chamber and opens the chamber connection to the electrometer. This causes a hard upscale transient on the x1 scale. Hold
Response Time
Approximately 5 s for 90% of final meter deflection on the x1 and x10 scales, and 3 s on the x100, x1k and x10k scales
Compensation
Automatically corrects for temperature and pressure changes in atmosphere
Pressure Compensation Range
70 to 106 kPa (20.7 to 31.3 in Hg)
Temperature Range
-20 to 50 °C (-4 to 122 °F), temperature compensation maintains calibration within 15% of 25 °C (77 °F) reading
Power
2 each “D” cell batteries housed in a sealed externally accessible compartment
Battery Life
400 hours