Model 15

Ludlum Model 15, neutron dose survey meter

Manufacturer
Ludlum Measurements
Reference
Model 15

Specifications

Product
Ludlum Model 15, neutron dose survey meter
Ludlum part number
48-1614
Indicated Use
Al pha, beta, gamma, neutron detection
Neutron Detector (Model 42-14H)
Detector Type: 2 atm 3 He proportional detector with 7.6 cm (3 in.) diameter, cadmium-lined (removable) moderator for thermal neutrons Sensitivity (bare 241 AmBe): 55 cpm per mrem/hr Detection Range: Thermal to approxima
Alpha-Beta-Gamma Detector (Model 44-7)
Detector Type: Thin end-window GM detector Window: 2.0 ± 0.3 mg/cm 2 mica Window Area: Active: 6.4 cm 2 (1 in 2 ); Open: 5.2 cm 2 (0.8 in 2 ) Efficiency (4π): 14 C: 2%, 90 Sr/ 90 Y: 10%, 99 Tc: 7%, 239 Pu: 7% Sensitivity
Instrument Range
0 to 500,000 cpm
Linearity
± 10% of true value
Meter Dial
0 - 500 cpm, 0 - 2.5 kV, BAT TEST (Meter Faces tab lists other available meter faces)
Meter
6.4 cm (2.5 in.) arc, pivot-and-jewel suspension
Controls
Multi-Selector Switch: Range multiplers (x1, x10, x100, x1000), Battery Check, Off Audio: On/Off switch Response Time: Pushbutton to zero meter Detector Switch: For displaying either the alpha-beta-gamma detector or neut
Audio
Built-in unimorph speaker with on/off switch, greater than 60 dB at 61 cm (2 ft.)
Working Environment
Normal working conditions
Temperature Range
-20 to 50 °C (-4 to 122 °F)
Power
2 each "D" cell batteries, housed in sealed compartment that is externally accessible
Battery Life
Approximately 600 hours with alkaline batteries, can be checked on meter
Connector
Series "C"
Size (H x W x L)
31.8 x 8.9 x 21.6 cm (12.5 x 3.5 x 8.5 in.)
Weight
3.7 kg (8.1 lb)
Meter Face
Image (click to view larger)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line