Model 375-31H

Ludlum Model 375-31H, area monitor with neutron detector

Manufacturer
Ludlum Measurements
Reference
Model 375-31H

Specifications

Product
Ludlum Model 375-31H, area monitor with neutron detector
Ludlum part number
48-4370
Indicated Use
Area monitoring, neutron detection
Detector
2 atm 3 He tube, LND 25185 or equivalent
Detector Moderator
22.9 cm (9 in.) diameter cadmium-loaded polyethylene sphere
Measurement Range
0 to 100 mSv/h (0 to 10 rem/hr)
Energy Response
Thermal to 12 MeV; approximately follows the radiation protection guide curve for neutron dose
Sensitivity ( 241 AmBe)
10,000 cpm per mSv/h (100 cpm per mrem/hr)
Gamma Rejection ( 137 Cs)
10 cpm or less, up to 0.1 Sv/h (10 R/hr)
Operating Voltage
1200 V (typical)
Input Sensitivity
-2 mV
Display
4-digit LED display with 2 cm (0.8 in.) character height
Display Range
000.0 to 9999 (Series One: 00.00 to 9999)
Display Units
µSv/h, mSv/h, mrem/hr, or rem/hr
Linearity
Readings within 10% of true value
Response
Typically 3 seconds from 10% to 90% of final reading
Indicators
STATUS: green light, instrument functioning properly LOW ALARM: yellow light and slow beep (1 per second), can be set at any point from 0.0 - 9999 HIGH ALARM: red light and fast beep (4 per second), can be set at any poi
Remote (optional)
Allows for connection of Ludlum Model 271 or 272 remote units
Relays
Fail-safe form C relays included
Connector
Series "C" (others available)
Ethernet (optional)
10 Base-T connection for use with Ludlum software
Calibration Controls
Accessible from the front of instrument (protective cover provided)