Model 3007

Ludlum Model 3007, neutron dose survey meter

Manufacturer
Ludlum Measurements
Reference
Model 3007

Specifications

Product
Ludlum Model 3007, neutron dose survey meter
Ludlum part number
m3007-series
Measurement Range
0 - 100 mSv/h (0 - 10,000 mrem/hr)
Detector
3 He proportional detector, 1.6 x 2.5 cm (0.6 x 1.0 in.) [D x L], surrounded by 19.6 cm (7.7 in.) diameter high density polyethylene sphere
Sensitivity
See Configurations tab
Energy Response
Provides an approximate inverse RPG curve for neutrons from thermal to 12 MeV
Gamma Rejection
Less than 10 cpm through 0.1 Sv/h (10 R/hr) [ 137 Cs gamma]
Threshold
-2 mV
High Voltage
Adjustable from 300 to 2400 Vdc
Operating Voltage
Approximately 1200 Vdc
Internal Detector ("i" versions)
Energy compensated GM or scintillation detector (see Configurations tab)
Linearity
Reading within 10% of true value
Modes
Four operational modes including rate, max, integrated dose, and count
Display
3-digit LCD with large 20 mm (0.8 in.) digits. Units: (k)(M)c, (k)cps, (k)(M)cpm, (k)(M)d, (k)(M)Bq, (k)(M)Bq/cm 2 , (k)(M)dpm, (μ)(m)(k)(M)R, (μ)(m)(k)R/h, (μ)(m)(k)(M)Sv, (μ)(m)Sv/h, (μ)(m)(k)(M)Rem, (μ)(m)(k)Rem/h Ind
Display Range
Minimum and maximum display can be set to limit display to calibrated range 0 cps to 999 kcps 0 cpm to 99.9 Mcpm 0 Bq to 999 MBq 0 Bq/cm 2 to 999 MBq/cm 2 0 dpm to 999 Mdpm 0 µR to 999 MR 0 µR/h to 9.99 kR/h 0 µSv to 999
Backlight
Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-on (will reduce battery life)
User Controls
ON/OFF (non-"i" versions): short press to turn ON, Short press to acknowledge alarms, Long press to reset RATE/MAX, Extra-long press plus 3 seconds to turn OFF ON/AUD ("i" versions): short press to turn ON, Short press t
Response Time
User-selectable from 1 to 60 seconds, auto SLOW, auto FAST, fixed SLOW (22 s), or fixed FAST (4 s)
Alarms
Three adjustable alarm setpoints for each unit in all four modes of operation
Self-Diagnostics
Detects and reports internal instrument issues, indicated by showing FAULT on the display and an audio warning
Overload Protection
High count rate saturation protection prevents false display of lower count rates. If the instrument detects an overload condition, it will show OVERLOAD on the display and trigger an audio warning.
Loss of Count Protection
After a user-settable number of seconds with no pulses from the detector, the instrument will trigger the alarm audio, and show LOSS OF COUNTS and the affected unit on the display. Disabled by default.