Model 30-7
Ludlum Model 30-7, neutron meters
- Manufacturer
- Ludlum Measurements
- Reference
- Model 30-7
- Product line
- Neutron survey meters
Specifications
- Product
- Ludlum Model 30-7, neutron meters
- Ludlum part number
- 30-7-series
- Indicated Use
- Neutron dose rate
- Detector
- 3 He proportional detector, 1.6 x 2.5 cm (0.6 x 1.0 in.) (D x L), surrounded by a 19.5 cm (7.7 in.) diameter polyethylene sphere with internal borated layer
- Measurement Range
- 0 to 99.9 mSv/h (0 to 9.99 rem/hr)
- Sensitivity (bare AmBe neutrons)
- Model 30-7: Typically 10 cpm per μSv/h (100 cpm per mrem/hr) Model 30-7B: Typically 4.5 cpm per μSv/h (45 cpm per mrem/hr)
- Energy Response
- Provides approximate inverse RPG curve for neutrons from thermal through 7 MeV, provides response up to 12 MeV (see Energy Response Curve on Graphs tab). Model 30-7 tends to overrespond in the 5 keV range.
- Linearity
- Reading within 10% of actual value
- Gamma Rejection ( 137 Cs gamma)
- Less than 10 cpm through 0.1 Sv/hr (10 R/hr)
- LCD Display
- 3 digit LCD with large 13.4 mm (0.53 in.) digits, low-battery indicator, MAX, ALARM, MUTE Units: (k)cps, (k)cpm, (µ)(m)Sv(/h), (µ)(m)rem(/h)
- Display Range
- 0.0 cps to 99.9 kcps 0.00 cpm to 999 kcpm 0.00 µSv/h to 999 Sv/h 0.00 µrem/hr to 999 rem/hr Display range can be set to limit display to calibrated range
- Backlight
- Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-On (will reduce battery life). Alarm light intensity varies based on ambient light levels.
- Controls
- ON/OFF/ACK: Long press to turn ON; Tap to acknowledge alarms and silence alarm tone; Press to reset Sigma Audio alarm, turn “click” audio On/Off, turn Sigma Audio beep On/Off; Press and hold for OFF UNITS: Long press cha
- Response Time
- User-selectable from 1 to 60 seconds, Auto-Response Rate FAST or SLOW, or Fixed FAST and SLOW
- Alarms
- Count rate, exposure/dose, and scaler alarm setpoints adjustable over the display range
- Overload
- High count rate saturation protection prevents false display of lower count rates
- Zero Protection
- After a user-settable time interval (default 60 seconds) of no pulses from the detector, the instrument will flash zero reading and the alarm audio will be triggered
- Dead Time Correction
- Employs first and second order corrections for extended performance
- Operating Voltage
- Approximately 1200 Vdc
- Threshold
- -2 mV
- Audio
- Greater than 75 dB at 0.6 (2 ft), approximately 4 kHz
- Click Audio Divide
- x1, x10, x100, x1k
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 12-4Ludlum Model 12-4, neutron dose survey meter
- Model 12-4-7Ludlum Model 12-4-7, neutron dose survey meter
- Model 15Ludlum Model 15, neutron dose survey meter
- Model 2241-4Ludlum Model 2241-4, neutron dose survey meter
- Model 2363Ludlum Model 2363, with model 42-41l neutron dose survey meter
- Model 30-4Ludlum Model 30-4, digital neutron survey meter
- Model 3004Ludlum Model 3004, neutron dose survey meter
- Model 3007Ludlum Model 3007, neutron dose survey meter
- Model 3014Ludlum Model 3014, with model 42-41l neutron dose survey meter
- Model 375-31HLudlum Model 375-31H, area monitor with neutron detector