Model 30-7

Ludlum Model 30-7, neutron meters

Manufacturer
Ludlum Measurements
Reference
Model 30-7

Specifications

Product
Ludlum Model 30-7, neutron meters
Ludlum part number
30-7-series
Indicated Use
Neutron dose rate
Detector
3 He proportional detector, 1.6 x 2.5 cm (0.6 x 1.0 in.) (D x L), surrounded by a 19.5 cm (7.7 in.) diameter polyethylene sphere with internal borated layer
Measurement Range
0 to 99.9 mSv/h (0 to 9.99 rem/hr)
Sensitivity (bare AmBe neutrons)
Model 30-7: Typically 10 cpm per μSv/h (100 cpm per mrem/hr) Model 30-7B: Typically 4.5 cpm per μSv/h (45 cpm per mrem/hr)
Energy Response
Provides approximate inverse RPG curve for neutrons from thermal through 7 MeV, provides response up to 12 MeV (see Energy Response Curve on Graphs tab). Model 30-7 tends to overrespond in the 5 keV range.
Linearity
Reading within 10% of actual value
Gamma Rejection ( 137 Cs gamma)
Less than 10 cpm through 0.1 Sv/hr (10 R/hr)
LCD Display
3 digit LCD with large 13.4 mm (0.53 in.) digits, low-battery indicator, MAX, ALARM, MUTE Units: (k)cps, (k)cpm, (µ)(m)Sv(/h), (µ)(m)rem(/h)
Display Range
0.0 cps to 99.9 kcps 0.00 cpm to 999 kcpm 0.00 µSv/h to 999 Sv/h 0.00 µrem/hr to 999 rem/hr Display range can be set to limit display to calibrated range
Backlight
Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-On (will reduce battery life). Alarm light intensity varies based on ambient light levels.
Controls
ON/OFF/ACK: Long press to turn ON; Tap to acknowledge alarms and silence alarm tone; Press to reset Sigma Audio alarm, turn “click” audio On/Off, turn Sigma Audio beep On/Off; Press and hold for OFF UNITS: Long press cha
Response Time
User-selectable from 1 to 60 seconds, Auto-Response Rate FAST or SLOW, or Fixed FAST and SLOW
Alarms
Count rate, exposure/dose, and scaler alarm setpoints adjustable over the display range
Overload
High count rate saturation protection prevents false display of lower count rates
Zero Protection
After a user-settable time interval (default 60 seconds) of no pulses from the detector, the instrument will flash zero reading and the alarm audio will be triggered
Dead Time Correction
Employs first and second order corrections for extended performance
Operating Voltage
Approximately 1200 Vdc
Threshold
-2 mV
Audio
Greater than 75 dB at 0.6 (2 ft), approximately 4 kHz
Click Audio Divide
x1, x10, x100, x1k