Model 2363
Ludlum Model 2363, with model 42-41l neutron dose survey meter
- Manufacturer
- Ludlum Measurements
- Reference
- Model 2363
- Product line
- Neutron survey meters
Specifications
- Product
- Ludlum Model 2363, with model 42-41l neutron dose survey meter
- Ludlum part number
- 48-3514
- Indicated Use
- Gamma and neutron survey
- Detection Range
- Model 42-41L PRESCILA Neutron Detector (External): 1 μSv/h to 10 mSv/h (0.1 mrem/hr to 1 rem/h) Energy-Compensated Gamma GM Detector (Internal): 1 μSv/h to 10 mSv/h (0.1 mR/hr to 1 R/hr)
- Sensitivity
- Neutron (with Model 42-41L): 35 cpm/μSv/h (350 cpm/mrem/hr), gamma rejection up to 1 mSv/h (100 mR/hr) Gamma (Internal Detector): 100 cpm/μSv/h (1000 cpm/mR/hr)
- Neutron Energy Response
- Thermal to 100 MeV
- Linearity
- Reading within 10% of true value with detector connected
- Gamma Rejection ( 137 Cs)
- Approximately 40 cpm at 1 mSv/h (400 cpm at 100 mR/hr)
- Meter Dial
- 0 - 100 μSv/h (0 - 10 mrem/hr), BAT OK
- Meter
- 6.4 cm (2.5 in.) arc, 1 mA analog type
- Digital Display
- 6-digit LCD with 0.64 cm (0.25 in.) digits
- Multipliers
- x0.1, x1, x10, x100
- Selector Switch
- Toggle switch to select gamma + neutron, gamma only, or neutron only
- Reset/Read ID
- A two-position momentary action switch that may be moved to select RESET the meter, or toward the READ ID position to read the integrated dose on the digital display
- Logging Pushbutton
- Located in the handle, used to store a reading
- Data Logger
- Capable of logging up to 1000 individual data points with the following identifiers for each point: Gamma and neutron sample counts Sample number Date/Time stamp Current integrated dose 10-character location identifier (
- Adjustable Alarms
- Gamma alarm Neutron alarm Integrated dose alarm (Indicated by red front panel LEDs)
- Response
- Will vary according to the number of counts present, typically 2 to 11 s from 10% to 90% of final reading
- Audio
- Built-in unimorph speaker with volume control (greater than 60 dB at 61 cm [2 ft], full volume)
- Audio Divide
- Selectable dual or individual click-per-event for gamma and neutron counts
- Power
- 2 each “D” cell batteries (housed in sealed compartment that is accessible from front of instrument)
- Battery Life
- Greater than 200 hours (battery condition can be checked on meter)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 12-4Ludlum Model 12-4, neutron dose survey meter
- Model 12-4-7Ludlum Model 12-4-7, neutron dose survey meter
- Model 15Ludlum Model 15, neutron dose survey meter
- Model 2241-4Ludlum Model 2241-4, neutron dose survey meter
- Model 30-4Ludlum Model 30-4, digital neutron survey meter
- Model 30-7Ludlum Model 30-7, neutron meters
- Model 3004Ludlum Model 3004, neutron dose survey meter
- Model 3007Ludlum Model 3007, neutron dose survey meter
- Model 3014Ludlum Model 3014, with model 42-41l neutron dose survey meter
- Model 375-31HLudlum Model 375-31H, area monitor with neutron detector