Model 3004

Ludlum Model 3004, neutron dose survey meter

Manufacturer
Ludlum Measurements
Reference
Model 3004

Specifications

Product
Ludlum Model 3004, neutron dose survey meter
Ludlum part number
m3004-series
Measurement Range
0-100 mSv/h (0-10,000 mrem/hr)
Detector
3 He proportional detector, 1.6 x 2.5 cm (0.6 x 1.0 in.) [D x L], surrounded by 22.9 cm (9.0 in.) diameter cadmium-loaded polyethylene sphere
Sensitivity
Typically 10 cpm per μSv/h (100 cpm per mrem/hr) [bare AmBe neutrons]
Energy Response
Provides an appropriate inverse RPG curve for neutrons from thermal through 7 MeV, provides response up to 12 MeV
Gamma Rejection
Less than 10 cpm through 0.1 Sv/h (10 R/hr) [ 137 Cs gamma]
Threshold
-2 mV
High Voltage
Adjustable from 300 to 2400 Vdc
Operating Voltage
Approximately 1200 Vdc
Linearity
Reading within 10% of true value
Internal Detector ("i" versions)
Energy compensated GM or scintillation detector (see Configurations )
Modes
Four operational modes including rate, max, integrated dose, and count
Display
3-digit LCD with large 20 mm (0.8 in.) digits. Units: (k)(M)c, (k)cps, (k)(M)cpm, (k)(M)d, (k)(M)Bq, (k)(M)Bq/cm2, (k)(M)dpm, (μ)(m)(k)(M)R, (μ)(m)(k)R/h, (μ)(m)(k)(M)Sv, (μ)(m)Sv/h, (μ)(m)(k)(M)Rem, (μ)(m)(k)Rem/h Indic
Display Range
Minimum and maximum display can be set to limit display to calibrated range. 00.0 cps to 999 kcps 00.0 cpm to 99.9 Mcpm 00.0 Bq to 999 MBq 00.0 Bq/cm² to 999 MBq/cm² 00.0 dpm to 999 Mdpm 0.00 μR to 999 MR 0.00 μR/h to 9.
Backlight
Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-on (will reduce battery life)
User Controls
ON/OFF (non-”i” version): press to turn ON, short press to acknowledge alarms and silence alarm tone, long press to reset RATE/MAX, hold for OFF ON/AUD (“i” versions): extra-long press to turn ON, short press to adjust a
Response Time
User-selectable from 1 to 60 seconds, auto SLOW, auto FAST, fixed SLOW (22 s), fixed FAST (4 s)
Alarms
Three adjustable alarm setpoints for each unit in all four modes of operation
Self-Diagnostics
Detects and reports internal instrument issues, indicated by showing FAULT on the display and an audio warning
Overload Protection
High count rate saturation protection prevents false display of lower count rates. If the instrument detects an overload condition, it will display the OVERLOAD icon and trigger an audio warning.
Loss of Count Protection
After typically 60 seconds of no pulses from the detector, the alarm audio will be triggered and the LOSS OF COUNTS icon and the affected unit will be displayed