Model 3014

Ludlum Model 3014, with model 42-41l neutron dose survey meter

Manufacturer
Ludlum Measurements
Reference
Model 3014

Specifications

Product
Ludlum Model 3014, with model 42-41l neutron dose survey meter
Ludlum part number
48-4507
External Detector
Model 42-41L PRESCILA neutron detector Sensitivity ( 241 AmBe): 35 cpm per µSv/h (350 cpm per mrem/h) Neutron Energy Response: thermal to 100 MeV neutrons Gamma Rejection ( 137 Cs): approx. 400 cpm at 100 mR/hr
Internal Detector
Energy compensated GM detector Sensitivity ( 137 Cs): 100 cpm per µSv/h (1000 cpm per mR/hr)
Measurement Range
Neutron: 1 µSv/h to 10 mSv/h (0.1 mrem/hr to 1 rem/hr) Gamma: 1 µSv/h to 50 mSv/h (0.1 mR/hr to 5 R/hr)
High Voltage
Neutron: typically 500 to 700 Vdc Gamma: 550 Vdc
Linearity
Readings within 10% of true value
Modes
Four operational modes including rate, max, count, and integrated dose
Display
3-digit auto-ranging LCD with large 20 mm (0.8 in.) digits. Units: (k)(M)c, (k)cps, (k)(M)cpm, (k)(M)d, (k)(M)Bq, (k)(M)Bq/cm2, (k)(M)dpm, (μ)(m)(k)(M)R, (μ)(m)(k)R/h, (μ)(m)(k)(M)Sv, (μ)(m)Sv/h, (μ)(m)(k)(M)Rem, (μ)(m)(
Display Range
00.0 cps to 999 kcps 00.0 cpm to 99.9 Mcpm 00.0 Bq to 999 MBq 00.0 Bq/cm 2 to 999 MBq/cm 2 00.0 dpm to 999 Mdpm 0.00 µR to 999 MR 0.00 µR/h to 9.99 kR/h 0.00 µSv to 999 MSv 0.00 µSv/h to 999 Sv/h 0.00 µRem to 999 MRem 0.
Backlight
Built-in ambient light sensor automatically activates LED backlight unless internal dipswitch is set to continuous-on
Controls
ON/AUD: Extra-long press to turn ON; Short press to adjust audio level; Extra-long press to turn OFF UNITS: Short press changes the units between count rate, dose/exposure, or disintegration DETECTOR: Short press to chan
Data Logging
Can store up to 1000 data points internally. Requires Lumic Data Logger Kit (see Options)
Response Time
User-selectable from 1 to 60 seconds, auto SLOW, auto FAST, fixed SLOW (22 s), fixed FAST (4 s)
Alarms
Three count rate, exposure/dose, count, and accumulated dose alarm set-points adjustable over the display range
Self-Diagnostics
Detects and reports internal instrument issues, indicated by graphic FAULT display and audio warning
Overload Protection
High rate saturation protection prevents false display of lower count rates. OVERLOAD will appear on the display and an audio warning will be triggered.
Loss of Count Protection
After a user defined time, typically 60 seconds, with no pulses from detector, the unit will flash a zero reading, an audio warning will be triggered, and LOSS OF COUNTS will appear on the display
Overrange
If reading should exceed the predefined detector range, the unit will flash a maximum reading, an audio warning will be triggered, and OVERRANGE will appear on the display
Audio
"Click" audio greater than 65 dB at 0.6 m (2 ft.), Alarm audio greater than 72 dB at 0.6 m (2 ft), approximately 4.5 kHz
Calibration Protection
Internal dipswitch to disable front-panel calibration parameter changes
Power
Four "AA" batteries