Model 9DP

Ludlum Model 9DP, pressurized ion chamber

Manufacturer
Ludlum Measurements
Reference
Model 9DP

Specifications

Product
Ludlum Model 9DP, pressurized ion chamber
Ludlum part number
48-3742
Radiation Detected
Gamma & X-rays above 25 keV Beta above 1 MeV
Energy Response
± 25% from 60 keV to 1.25 MeV (see Graph )
Display Ranges (auto-ranging)
Sv/h units *: 0 - 5 μSv/h, 0 - 50 μSv/h, 0 - 500 μSv/h, 0 - 5 mSv/h, 0 - 50 mSv/h R/hr units: 0 - 500 μR/hr, 0 - 5 mR/hr, 0 - 50 mR/hr, 0 - 500 mR/hr, 0 - 5 R/hr Gy/h units: 0 - 5 μGy/h, 0 - 50 μGy/h, 0 - 500 μGy/h, 0 -
Measurement Readouts
Simultaneous display of rate and either the integrated value or highest (peak) rate
Minimum Display
0.01 µSv/h (0.1 µR/hr, 0.01 µGy/h, 0.1 µrem/h)
Drift
Less than 0.3 µSv/h (0.03 mR/hr, 0.3 µGy/h, 0.03 mrem/h)
Chamber Volume
230 cc (14 in 3 ) volume pressurized to 9 atm (117 psig ± 5 psig)
Chamber Density
Chamber wall density is 601.7 mg/cm 2 , Can wall density is 332.5 mg/cm 2 . Total density of chamber + can is 934.2 mg/cm 2 .
Accuracy
± 10%
Response Time
Ranges from 7 seconds in lowest range to under 2 seconds in highest range when measuring from 10% to 90% of final value
Display
LCD with 8.9 cm (3.5 in.) diagonal, 240 H x 320 W pixels, TFT active matrix, 232K colors, 220 cd/m 2 , automatic sensor-controlled backlighting
User Controls
4 pushbuttons: ON/OFF FUNCTION (for peak rate / integrate modes) AUDIO (on/off, volume) ACK/RESET (for alarm acknowledgement, meter reset, and clearing integrated exposure or peak rate)
Automatic Functions
Auto ranging, Auto zeroing, Auto LCD backlighting
Alarms
Two levels of radiation alarms available, each is user programmable throughout entire readout range.
Data Logging
Data is stored to detachable USB drive in CSV format for easy retrieval by PC spreadsheet/database programs. Data points include date and time, rate, integrated reading, and instrument status. Logging time intervals are
Audio Outputs
Built-in unimorph speaker, > 60 dB at 0.6 meters (2 ft.) Optional audio jack available for connection to external headset. (see Options )
USB Interface
Single USB 2.0 port. May be connected to: A USB keyboard (with no additional USB ports, and no integrated mouse or trackpad or audio controls) to facilitate password-protected parameter changes A USB memory storage devic
Warm Up Time
1 minute when the instrument is in temperature equilibrium with the surrounding environment
Temperature Range
-20 to 40 °C (-4 to 104 °F)
Humidity
0 - 95%, non-condensing

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item